Installation, Operating & Maintenance Instructions
Series 650 DN 100-
250 (I.D. 4“ - 10”), DeviceNet
VAT Vakuumventile AG, CH-9469 Haag, Switzerland
Tel +41 81 771 61 61 Fax +41 81 771 48 30 [email protected] www.vatvalve.com
266341ED
2012-07-13
31/79
3.6.2 LOGIC I/O configuration
The «logic digital input close valve» can adjusted. See also chapter «LOGIC I/O connection».
The «logic digital input close valve»
can be configured by „Control View‟, „Control Performance Analyzer‟or „Universal Hyper
Terminal.
The
«set-command»
is:
«s:20000000a0»
.
The
«get-command»
is:
«i:20»
. The answer returns the «operational settings mode»
«i:20000000a0».
Description:
code
description
data range
s:20
command header
a
mode selection
0 = non inverted
1 = inverted
2 = disabled
3.6.3 Valve and sensor configuration
Basic valve configuration must be adapted according to application needs.
Definition of valve plate position (CLOSE or OPEN) after power up sequence. Default is „close„.
Definition of valve
plate position (CLOSE or OPEN) in case of a power failure. Default is „not defined„.
Only for versions that have Power Fail Option equipped [650 . . - . .
C
. - . . . . or 650 . . - . .
H
. - . . . . ].
ZERO function. This may be „disabled‟ or „enabled‟. Default is „enabled„. Refer also to «ZERO (setup step 4)».
Sensor configuration for 2 sensor version [650 . . - . . .
Q
- . . . .].
Refer also to «Pressure control operation with 2 sensors».
Local operation
:
(„Control View‟, „Control Performance Analyzer‟ or
„Service Box 2„)
Remote operation
:
(Refer to chapter «Explicit messaging setup
commands» for details)
1. Do power up configuration in menu
„Setup / Valve‟.
1. Select POWER UP CONFIGURATION
2. Do power fail configuration in menu
„Setup / Valve‟.
2. Select POWER FAIL CONFIGURATION
3. Enable or disable ZERO function in menu
„Setup / Sensor‟.
3. Select ZERO CONTROL
4. Do 2 sensor configuration in menu
„Setup / Sensor‟.
4. Select SENSOR MODE
5. Select SENSOR RATIO