33
20901220_EN_VACUU·PURE 10_V1.6_060722
Installation and connection
Set up the vacuum pump
5 cm
12 cm
5 cm
>
Place the vacuum pump on a stable, non‑vibrating, level
surface.
>
When installing in lab furniture, maintain a minimum
distance of 5 cm (2 in) to adjacent objects or surfaces.
>
The product must be positioned so that the on/off switch
and the power plug can be reached and are accessible,
minimum distance 12 cm (5 in).
>
Prevent heat accumulation and ensure sufficient air
circulation, especially in closed housings.
>
Always ensure that there is an adequate air supply and air
extraction to remove the warm exhaust air from the vacuum
pump. Provide external forced ventilation with a volume flow
of approx. 100 m
3
/h when installing in lab furniture.
Observe limitation of use
Limitation of use
(US)
Operating ambient temperature
10 – 40 °C
50 – 104 °F
Max. altitude
2000 m
above sea level
6562 ft
above sea level
Minimum distance to adjacent
parts
5 cm (12 cm)
2 in (5 in)
Relative humidity
30
–
85 %, non‑condensing
Pollution degree
2
Protection class
IP 20
NEMA type 1
Prevent condensation or external contamination from dust, liquids, and
corrosive gases.
IMPORTANT!
Example
Sketch of
minimum
distances in lab
furniture
Observe limitation
of use
Содержание VACUU-PURE 10
Страница 1: ...Instructions for use OI no 20901220 Original instructions EN VACUU PURE 10 Vacuum pump...
Страница 12: ...12 20901220_EN_VACUU PURE 10_V1 6_060722 Introduction...
Страница 30: ...30 20901220_EN_VACUU PURE 10_V1 6_060722 Safety information...
Страница 83: ...83 20901220_EN_VACUU PURE 10_V1 6_060722 Appendix 8 7 8 7 CU Certificate CU Certificate CU Certificate...
Страница 84: ...84 20901220_EN_VACUU PURE 10_V1 6_060722...
Страница 85: ...85 20901220_EN_VACUU PURE 10_V1 6_060722...
Страница 86: ...86 20901220_EN_VACUU PURE 10_V1 6_060722...
Страница 87: ...87 20901220_EN_VACUU PURE 10_V1 6_060722...