61
20901220_EN_VACUU·PURE 10_V1.6_060722
Error messages
Error
Error
`
Possible cause
3
Remedy
Personnel
Error
flashing rate
6x
`
VACUU·BUS
accessories removed
/ unplugged.
3
Plug in
VACUU·BUS
accessories again
and switch vacuum
pump off/on.
3
Operation without
VACUU·BUS
accessories:
Perform bus scan.
Operator
`
Error or
communication
interruption in
VACUU·BUS
accessories.
3
Check VACUU·BUS
plug‑in connection
to accessories.
3
Replace defective
components.
Error
flashing rate
7x
`
Other errors (e.g.,
incompatible software
version, other
frequency converter
errors).
3
Perform or repeat
the software
update. Information
about software
updates:
3
Send in vacuum
pump.
Resp.
specialist
Optional
accessories:
Vacuum
sensor does
not display
a measured
value.
`
No voltage applied.
3
Apply mains
voltage, switch on
vacuum pump.
Operator
`
VACUU·BUS plug‑in
connection or cables
defective or not
connected.
3
Check VACUU·BUS
plug‑in connection
and wiring.
`
External plug‑in
power supply of
vacuum sensor not
plugged in.
3
Plug‑in power
supply for vacuum
sensor.
`
Sensor defective.
3
Replace defective
components.
Specialist
Error – Cause –
Remedy
Содержание VACUU-PURE 10
Страница 1: ...Instructions for use OI no 20901220 Original instructions EN VACUU PURE 10 Vacuum pump...
Страница 12: ...12 20901220_EN_VACUU PURE 10_V1 6_060722 Introduction...
Страница 30: ...30 20901220_EN_VACUU PURE 10_V1 6_060722 Safety information...
Страница 83: ...83 20901220_EN_VACUU PURE 10_V1 6_060722 Appendix 8 7 8 7 CU Certificate CU Certificate CU Certificate...
Страница 84: ...84 20901220_EN_VACUU PURE 10_V1 6_060722...
Страница 85: ...85 20901220_EN_VACUU PURE 10_V1 6_060722...
Страница 86: ...86 20901220_EN_VACUU PURE 10_V1 6_060722...
Страница 87: ...87 20901220_EN_VACUU PURE 10_V1 6_060722...