56
20901220_EN_VACUU·PURE 10_V1.6_060722
Error messages
6
6 Error messages
Error messages
Errors or warnings are indicated by the colored LEDs on the
warning triangle. Several error messages may be pending at the
same time. Errors and warnings can be identified via the flashing
rate.
LED
Description
Gray
No warning or error active
Yellow
Warning
Warning messages reset themselves as soon as the values
are back within the normal range. The vacuum pump
continues to run if a warning message is received.
Red
Error
The vacuum pump stops as soon as there is an error. If
there is an error, all warnings are ignored. First correct the
error before resetting.
6.1
6.1 Warning display
Warning display
Flashing
Description
1x
Temperature in critical range
2x
Not assigned
3x
Motor current consumption in critical range
4x
Fan speed deviation
5x
Supply voltage for control board in critical range
6x
Messages: VACUU·BUS accessories (e.g.,
overpressure of a vacuum sensor)
7x
Other warnings
General error
messages
Possible flashing
rates in the event of
a warning
Содержание VACUU-PURE 10
Страница 1: ...Instructions for use OI no 20901220 Original instructions EN VACUU PURE 10 Vacuum pump...
Страница 12: ...12 20901220_EN_VACUU PURE 10_V1 6_060722 Introduction...
Страница 30: ...30 20901220_EN_VACUU PURE 10_V1 6_060722 Safety information...
Страница 83: ...83 20901220_EN_VACUU PURE 10_V1 6_060722 Appendix 8 7 8 7 CU Certificate CU Certificate CU Certificate...
Страница 84: ...84 20901220_EN_VACUU PURE 10_V1 6_060722...
Страница 85: ...85 20901220_EN_VACUU PURE 10_V1 6_060722...
Страница 86: ...86 20901220_EN_VACUU PURE 10_V1 6_060722...
Страница 87: ...87 20901220_EN_VACUU PURE 10_V1 6_060722...