62
20901220_EN_VACUU·PURE 10_V1.6_060722
Error messages
Error
Error
`
Possible cause
3
Remedy
Personnel
Vacuum
pump does
not start.
`
Vacuum pump
switched off.
3
Switch on vacuum
pump at toggle
switch.
Operator
`
Power plug not
correctly plugged in
or pulled out.
3
Check power
connection and
cable.
`
Overpressure in the
outlet line.
3
Open the outlet line.
`
External silencer
(optional) clogged or
jammed.
3
Clean or replace
external silencer.
3
Remove silencer
and connect outlet
line instead.
`
Motor overloaded.
3
Allow the motor to
cool down.
Resp.
specialist
`
Excessive
temperature error
flashing rate 1x.
3
See error, flashing
rate 1x.
`
Vacuum pump
mechanically
blocked.
3
Send in vacuum
pump.
Ultimate
vacuum not
reached.
`
Leak in inlet line or in
apparatus.
3
Check inlet line
and equipment for
leaks.
Operator
`
Vacuum pump
not at operating
temperature.
3
Allow vacuum pump
to warm up for 20
minutes with inlet
closed.
`
Leak inside vacuum
pump.
3
Send in vacuum
pump.
Resp.
specialist
Error – Cause –
Remedy
Содержание VACUU-PURE 10
Страница 1: ...Instructions for use OI no 20901220 Original instructions EN VACUU PURE 10 Vacuum pump...
Страница 12: ...12 20901220_EN_VACUU PURE 10_V1 6_060722 Introduction...
Страница 30: ...30 20901220_EN_VACUU PURE 10_V1 6_060722 Safety information...
Страница 83: ...83 20901220_EN_VACUU PURE 10_V1 6_060722 Appendix 8 7 8 7 CU Certificate CU Certificate CU Certificate...
Страница 84: ...84 20901220_EN_VACUU PURE 10_V1 6_060722...
Страница 85: ...85 20901220_EN_VACUU PURE 10_V1 6_060722...
Страница 86: ...86 20901220_EN_VACUU PURE 10_V1 6_060722...
Страница 87: ...87 20901220_EN_VACUU PURE 10_V1 6_060722...