55
20901220_EN_VACUU·PURE 10_V1.6_060722
Commissioning (operation)
(1)
4
4.
4.
Switch off rocker switch
(1)
– switch position
0
.
5
Vacuum pump switched off.
5.
5.
Disconnect the vacuum pump from the apparatus.
6.
6. Check the vacuum pump for possible damage and
contamination.
5.6
5.6 Storage
Storage
Store the vacuum pump
1.
1. Carry out the steps for decommissioning
see chapter:
Shutdown (switch off) on page 54
Shutdown (switch off) on page 54
.
2.
2.
Clean the vacuum pump in the event of external
contamination.
3.
3. Close the vacuum pump's inlet and outlet, e.g., with the
transport locks.
4.
4.
Package the vacuum pump such that it is protected from
dust; enclose desiccants if necessary.
5.
5. Store the vacuum pump in a cool, dry location.
If damaged parts are stored for operational reasons, these
should be clearly identified as not ready for use
not ready for use
.
Store the vacuum
pump
IMPORTANT!