18
20901220_EN_VACUU·PURE 10_V1.6_060722
Safety information
2.4.3
2.4.3 Laboratory and working materials
Laboratory and working materials
DANGER
DANGER
Hazardous substances could be discharged at
the outlet.
During aspiration, hazardous, toxic substances at the
outlet can get into the ambient air.
>
Observe the relevant safety regulations for safe
handling of hazardous substances.
>
Please note that residual process media may pose a
danger to people and the environment.
>
Mount and use suitable separators, filters or fume
hood devices.
>
Prevent the release of hazardous, toxic, explosive, corrosive
fluids, gases or vapors or those that are harmful to health or
the environment, for example, through suitable laboratory
facilities with a fume hood and ventilation control.
Hazards due to different substances
Pumping different substances or media can cause the substances
to react with one another.
>
Be aware of interactions and possible chemical reactions of
the pumped media.
>
Dry the vacuum pump with ambient air before changing
the medium to be pumped. Use regeneration mode of the
vacuum pump for this purpose
see chapter:
5.2.2 Regeneration mode on page 45
5.2.2 Regeneration mode on page 45
2.4.4
2.4.4 Chemical compatibility of materials
Chemical compatibility of materials
Working materials which get into the vacuum pump with the gas
flow can damage the vacuum pump. Substances can be deposited
in the vacuum pump.
>
Check the compatibility of the pumped substances with the
wetted materials of the vacuum pump
see chapter:
8.1.3 Wetted materials on page 75
8.1.3 Wetted materials on page 75
>
Contact us if you have concerns about using your vacuum
pump with certain working materials or media.
Pumping different
substances
Compatibility of the
vacuum pump with
pumped substances
Содержание VACUU-PURE 10
Страница 1: ...Instructions for use OI no 20901220 Original instructions EN VACUU PURE 10 Vacuum pump...
Страница 12: ...12 20901220_EN_VACUU PURE 10_V1 6_060722 Introduction...
Страница 30: ...30 20901220_EN_VACUU PURE 10_V1 6_060722 Safety information...
Страница 83: ...83 20901220_EN_VACUU PURE 10_V1 6_060722 Appendix 8 7 8 7 CU Certificate CU Certificate CU Certificate...
Страница 84: ...84 20901220_EN_VACUU PURE 10_V1 6_060722...
Страница 85: ...85 20901220_EN_VACUU PURE 10_V1 6_060722...
Страница 86: ...86 20901220_EN_VACUU PURE 10_V1 6_060722...
Страница 87: ...87 20901220_EN_VACUU PURE 10_V1 6_060722...