3.3.5 Temperature
Probe
The temperature probe is mounted in the bottom left of the Module C, in
the inlet tee. To install, make certain that the plug is inserted in the inlet
tee. Drop the cable out of the bottom of the Module C, and affix the
plastic disc in the hole. Plug the connector into the rear right of the
Controller. See the Wiring Diagram for more details.
3.3.6
(Optional) Collocated Module C
The optional Collocated Module C is installed exactly the same as the standard Module C. The only
differences is that (2) of the cables are much longer so that the Collocated Module C can be installed 1
meter (inlet to inlet) away from the Module C. See the wiring diagram below for details.
3.3.7
Bottom-View Wiring Diagram
The following diagram shows a bottom view of the modules for wiring. Cable lengths are not to scale.
These bottom views are as if the front of the Controller, Module C, or Collocated Module C are facing up.
Example
: Take a Controller with the front facing you and you lie it on it’s back. The bottom view that you
would see is what is detailed below.
T-Plug with
Temperature
Probe Cable
Air Line
Receptacle
Controller
Cable
Receptacle
Module C
(Bottom)
Co
C
Module C
Cable
Receptacle
Pump Relay
Cable
Receptacles
Controller
(Bottom)
Power
Receptacle
Temperature
Probe
Receptacles
Collocated
Module C Cable
Receptacle
T-Plug
Air Line
Receptacle
Collocated
Module C
(Bottom)
Controller
Cable
Receptacle
Air Line
Receptacle
Power
Cord
Pump Relay
Cable
Receptacles
Mass Flow
Controller
Receptacle
Collocated Module C
Pump Enclosure
(Side of Lower Stand)
Power
Cord
Air Line
Receptacle
Mass Flow
Controller
Receptacle
Pump Relay
Cable
Receptacles
Module C
Pump Enclosure
(Side of Lower Stand)
19