4.9.1.1 Site Configuration
Location Code
000000000
Pressing
↑
F1
at the fifth Main Menu begins the Site Configuration process.
33
The first screen of the Site Configurations prompts you for a Location
Code. After entering the proper code with the digit keys, press
ENTER
to
continue.
ENTER=Done
Serial Number
0000
The next screen prompts you for the URG 3000-N’s Serial Number. Press
ENTER
to continue.
ENTER=Done
The Number of Modules screen appears next, which allows you to select 1
or 2 modules (if the configuration contains a collocated Module C).
Number of Modules
1 or 2
Current: 1
Enter=Done
The next screen allows you to choose a sampling schedule. Here, the
options shown are 1 in 3 days, and 1 in 6. Currently selected schedule is
also displayed. Press
ENTER
continues to the next screen.
Select Schedule
1: 1 in 3 2: 1 in 6
Current: 1
ENTER=Done
Sampling Interval
Typ:15 Max:60 min
15 minutes
Sampling Interval is the screen that follows schedule. It allows you to
change the time in minutes between samples. As shown, this is typically 15
minutes, whereas the maximum is set at 60 minutes. Press
ENTER
to
continue
ENTER=Done
Sample Time Per Hour
Typ:60 Max:60 min
60 minutes
Sample Time Per Hour allows the choice of how long a sample will run
during each hour. This is typically set at 60 minutes, with a maximum of 60
minutes, to allow for shorter sampling hours if necessary. Press
ENTER
to
continue.
ENTER=Done
Filter Configuration
1=Normal 2=Stacked
Current: 1
ENTER=Done
The final site configuration screen allow for you to select from Normal or
Stacked Filter Configuration, as shown. When completed, press
ENTER
to
complete the Site configuration process, and return to the fifth Main Menu.
4.10 Collected Data
The URG-3000N saves text files to the root directory of a CompactFlash memory card when sampling or
other procedures have occurred. Example filenames include c0000105.299, a0000105.299 or
r0000106.008. Note that files beginning with c are calibration logs, a are audit logs and r is sample data.
An example of the headers from a sampling data logfile (with text wrap) is shown below. These headers
are different for calibration/audit/sample runs. Values would be listed below.
SN LOCATION_NUM Q_NUMBER COMP_ID_NUM SDATE STIME EDATE ETIME TempAV
TempMN TempMX BaroAV BaroMN BaroMX VacAV VacMN VacMX FlowAV FlowMN FlowMX RTmpAV RTmpMN RTmpMX
RBarAV RBarMN RBarMX RVacAV RVacMN RVacMX RFloAV RFloCV RFloVL GAIN OFFSET OPI ADC BF
VERSION