Cal pt: 1 Mod:[1]
Gain=6.00 Off= 0.00
Raw=2800 Flow= 19.77
43
You are now prompted to enter a Reference Flowrate in LPM. This
completes the first calibration point.
Enter Ref(LPM):?
Calib. Results Mod:1
Gain=0.000 Off=0
Correlation= 0.00
The Calibration Results screen will be and will allow you to save the flow
calibration.
Save? YES/NO
Calib. Results Mod:2
Gain=0.000 Off=0
Correlation= 0.00
The software will advance through the 2
nd
and 3
rd
calibration points exactly
the same way it did for the 1
st
calibration point.
SAVED
After entering the 2
nd
and 3
rd
calibration point, the Calibration Results
screen will be shown for each calibration point and will allow you to save
the flow calibration results for each point.
Calib. Results Mod:1
Gain=5.98 Off=.25
Correlation= 1.00
After choosing one of these options, the Calibration results will be
displayed again and selecting
ENTER
will return to the Calibration Menu.
ENTER=Done
.
6.4 Flow Calibration for Collocated Module C
The flow calibration procedure for the Collocated Module C is done exactly
the same way as the Module C. Follow instruction in Section 6.3 to do a
flow calibration on the Collocated Module C.
Calibration Mod:[2]
NO=Back ENTER=Next
Select
Mod: 2
for doing a flow calibration on the Collocated Module C, and
ENTER
to begin.