In the example on the right, Gain is being modified. Pressing
↑
F1
would
toggle b and – values, and the digit keys would be used to enter a
number, using
BKSP
key to make any corrections.
Manual Cal. Mod:[1]
F1: +/-
Gain: 0.000
NO=Cancel ENTER=Done
4.6.3.3 Display Temp, BP, Vacuum
Pressing the
←
F3
key at the Maintenance Menu brings you to the Display
Temperature, Barometric Pressure, and Vacuum screen. After a series of
warm-ups, this screen displays the current Vacuum data for any available
Modules, and shows the Barometric Pressure and Temperature reading
from the Module C. Pressing
ENTER
will return you to the Maintenance
Menu.
Display Temp,BP,Vac
Vac: 1= 22.0 2= 22.0
Temp: 25.7 BP: 761.3
ENTER=Done
4.6.3.4 Pumps On/Off
Pump Status
Mod:[1] Pump Off
Mod:[2] Pump Off
Pressing
↑
F1
at the second Maintenance Menu will allow you to access
the Pump Status screen. Pressing
1
or
2
will toggle On/Off for the
corresponding pump. Pressing
ENTER
will returns to the Maintenance
Menu.
12=Pump# ENTER=Back
4.6.3.5 Solenoids On/Off
Solenoid Status
Sol[1]:0 Sol[2]:0
Sol[3]:0 Sol[4]:0
Pressing
↓
F2
at the second Maintenance Menu will bring you to the
Solenoid Status screen. Pressing
1
,
2
,
3
, or
4
will turn the corresponding
solenoid valves On/Off. Pressing
ENTER
will return you to the
Maintenance Menu.
1234=Sol# ENTER=Back
4.6.4 Audit
Audit Mod:[1]
F1=Leak Ck F2=Flow
F3=Temp. F4=BP
30
Pressing
←
F3
on the second Main Menu screen brings you to the Audit
menu. The Audit menu allows you to choose between a Leak Check
procedure, a Flow Rate Audit, a Temperature Audit, and a Barometric
Pressure Audit. Pressing
ENTER
will return to the second Main Menu.
ENTER=Back
4.6.4.1 Leak Check
Leak Check Mod:[1]
NO=Cancel ENTER=Next
Pressing the
↑
F1
key at the Audit Menu will bring you to the Leak Check
procedure. Note that the selected module is located in the top right. Press
Key
1
or
2
to select between a Module C or Collocated Module C.
Press
ENTER
to continue the Leak Check, or press
NO
to cancel.