Section 6 Calibration
This menu allows you to perform the temperature, barometric pressure, and flow calibrations. The
temperature and barometric pressure calibrations are only done on the Module C. The flow calibration
should be performed on the Module C, and the Collocated Module C if one is present.
6.1 Temperature
Calibration
Raw Offset C F
1457 0 20.0 68.0
SPACE=Calibrate
41
Pressing
↑
F1
at the calibration menu will bring you to the Temperature
Calibration screen shown on the right. Refer to Section 4.12 for a
complete temperature calibration software menu map. It displays the
current DAC Raw
ENTER=Back
values, Offset, Temperature in Celsius, and Fahrenheit. Pressing
SPACE
allows you to proceed with temperature calibration.
Raw Offset C F
1457 0 20.0 68.0
F1:+/- F2:C/F
You are now able to enter a positive or negative (toggled by
↑
F1
)
temperature value using the digit keys Celsius or Fahrenheit (toggled by
↓
F2)
. Use the
BKSP
key to clear the currently entered values.
After pressing
ENTER
, you are shown the final calibration temperature, as
shown to the right. Pressing
YES
will save the data, pressing
NO
will
cancel the calibration. You will be returned to the Calibration Menu.
Ref. Temp):?
Calibration Temp:
20.0 degrees C
Raw=1457 Offset= 0
YES=Save NO=Cancel
Calibration Temp:
20.0 degrees C
Raw=1457 Offset= 0
The next screen will confirm that the temperature calibration results have
been saved to the memory card. Press
ENTER
to return to the Calibration
Menu.
SAVED
6.2 Barometric Pressure Calibration
Raw Offset BP
2753 0 639.4
SPACE=Calibrate
Pressing the
↓
F2
at the Calibration Menu will allow you to proceed with
BP calibration. Refer to Section 4.12 for a complete barometric pressure
calibration software menu map. Initially, the screen shown to the right will
display Raw, Offset, and BP values. Pressing
SPACE
will allow you to
proceed to the following screen.
The next screen allows you to enter a reference BP value in mmHg using
the digit keys.
↑
F1 /
↓
F2
will change between positive and negative values.
Use the
BKSP
key to clear incorrect data.
ENTER=Back
Raw Offset BP
2753 0 639.4
Ref. BP(mmHg):?