70
8
STANDARD INTERFACE
70
STANDARD INTERF
ACE
Chapter
8
◇
SI/F print setting method
1.
Select setting mode 2.
Press
→
once.
2.
Set "SI/F print."
Press
five times
⇒
"SI/F function select" is
displayed.
Press
, and the highest digit blinks. Select the digit
of the SI/F print.
Input the setting value with
.
(Initial value: 0)
Press
to validate it.
3.
Press
twice to go back to the indicated value
display.
-
Operation of the indicated value hold function
For the "Stable" parameter, see "5-8.Motion detect " on page 44, and for the "Near zero"
parameter, see "6-3.Near zero " on page 52.
When stable value print is selected, be aware that automatic printing is not
performed in the following case.
- When motion detect is set in time: 0.0 sec, count: 000.
Also, the indicated value is not held in the following case.
- Hold mode:
When any item other than none is selected.
Point
Speed
rpm
Angle
rad deg
FNC
Torque
Speed
rpm
Nm
0
0
0
《
Setting value
》
H I
LO
HOLD
SI/F print
0: None
Undefined
1: Stable value
Undefined
2: Hold value
Angle
rad deg
RECALL
RECALL
ESC
Indicated value
Near zero
0
Stable
Near zero
ON
ON
ON
ON
OFF
OFF
OFF
Automatic printing
↑
Hold
1.5seconds
Time
ON
*Hold function …
Indicated value
SI/F
Comparison output signal
+
Sensor input value
Near zero
0
Time
Stable
Near zero
ON
ON
ON
ON
OFF
OFF
OFF
Automatic printing
↑
Hold
1.5seconds
ON
-
Indicated
Sensor input value
value
The print command and the record function are closely related. For details, see
"9-1.Record timing " on page 77.
Point
Содержание TM380
Страница 1: ...TM380 20MAY2021REV 1 02 TORQUE MONITOR OPERATION MANUAL ...
Страница 11: ...Contents X Contents X M E M O ...
Страница 147: ......