CALIBRATION PROCEDURE
27
CALIBRATION PROCE
DURE
Chapter
4
⑦
a.Equivalent input calibration
Check the specification of the sensor, and register the rated output value and the value you want to
display at that time (rated capacity value).
Input the rated output value and display value (rated capacity value) successively.
Equivalent input calibration is not executed by simply inputting the rated output value.
- UTM
Ⅲ
≪
Setting value
≫
Rated output value [V]:
-10.000 to 10.000 (except 0)
Display value (rated capacity):
-99999 to 99999 (except 0)
- UTM
Ⅱ
≪
Setting value
≫
Rated output value [V]:
-5.000 to 5.000 (except 0)
Display value (rated capacity):
-99999 to 99999 (except 0)
By performing equivalent input calibration, the setting value of actual load calibration becomes the
input rated capacity value.
However, it stays unchanged when an error occurs.
◇
Equivalent input calibration setting method
1.
Select setting mode 4.
Press
→
three times.
2.
Set "Rated output value" of the sensor.
Press
three times.
⇒
"Equivalent input calibration"
is displayed.
Enter a setting value input state with
, and the
highest digit blinks.
Check the specification of the sensor in use, and input the
rated output value (V) with
.
Move to the lower digits with
. (Initial value: 10.000)
Press
to validate it.
3.
Set "Rated capacity value."
Decide the load you want to display at the rated output
value, and input it with
.
Move to the lower digits with
.
Press
to validate it.
* For inputting a negative sign, press
.
Set a display value (rated capacity value) converted according to the set unit and
decimal place of torque.
For unit conversion, see "4-7.Unit conversion table" on page 39.
Point
Speed
rpm
Angle
rad deg
FNC
Torque
Speed
rpm
Nm
H I
LO
HOLD
Angle
rad deg
RECALL
RECALL
Torque
Speed
rpm
Nm
H I
LO
HOLD
Angle
rad deg
RECALL
+/-