47
5
EXPLANATION OF INDICATED- VALUE-RELATED FUNCTIONS
47
EXP
LANATION OF
INDICATED-
VALUE-RELATED FUNCTIONS
Chapter
5
5-10.
Display frequency
Set the per-second display update frequency of the indicated value. However, the A/D conversion
speed does not change.
≪
Setting value
≫
0: 3 1: 6 2: 13 3: 25 [times/sec.]
■
Display frequency setting method
1.
Select setting mode 2.
Press
→
once.
2.
Set "Display frequency."
Press
six times
⇒
"Display frequency" is displayed.
Press
, and present set value blinks.
Input the setting value with
.
Press
to validate it. (Initial value: 3)
3.
Press
twice to go back to the indicated value display.
CAUTION
●
Zero tracking is equal to digital zero functioning automatically. Therefore, the digital
zero limit is also monitored.
●
Zero tracking does not function if the indicated value has already exceeded the
tracking band.
- When displacement of the zero point is within the set count of tracking and it
continues more than the set time, it is automatically made zero by zero tracking
function.
- The time (tracking delay) is set in the range of 0.1 - 9.9 sec., and the band
(tracking band) is set in the range of 001 to 999.
If the time is set at 0.0 sec. and the band at 000, the zero tracking function does
not work.
Point
Boundary of zero track
DELAY
Indicated value
+COUNT
-COUNT
BAND
+
0
-
DELAY
Band=count×2
From the point when it returned within the range,
counting will be resumed.
Speed
rpm
Angle
rad deg
FNC
Torque
Speed
rpm
Nm
H I
LO
HOLD
Angle
rad deg