Chapter
10
■
Data select
When the output mode is "0: Data select", the analog output for torque, rotation speed and angle
will be synchronized with the indicated value.
The output data for each indicated value can be selected.
《
Setting value
》
0: Hold synchronized
Outputs value synchronized with the indicated value.
1: Hold not synchronized
Outputs value without hold even when the indicated value is on
hold.
◇
Output mode/ data select setting method
1.
Select setting mode 5.
Press
→
four times.
2.
Set " Output data select."
Press
once.
⇒
"Output data select" is
displayed.
Press
, and the highest digit blinks.
①
Set "Output mode."
Input the output mode with
,
and move to the lower digit with
.
(Initial value: 0)
②
Set "Data select 3 (angle)."
Input the data select 3 (angle) with
,
and move to the lower digit with
.
(Initial value: 0)
③
Set "Data select 2 (rotation speed)."
Input the data select 2 (rotation speed) with
, and move to the lower digit
with
. (Initial value: 0)
④
Set "Data select 1 (torque)."
Input the data select 1 (torque) with
. (Initial value: 0)
Press
to validate it.
3.
Press
twice to go back to the indicated value display.
Speed
rpm
Angle
rad deg
FNC
《
Setting value
》
0
0
0
0
Data select 1
0: Hold synchronized
1: Hold not synchronized
(torque)
Data select 2
0: Hold synchronized
1: Hold not synchronized
(rotation speed)
Data select 3
0: Hold synchronized
1: Hold not synchronized
(angle)
Output mode
1: Zero scale fix
2: Full scale fix
0: Data select
Torque
Speed
rpm
Nm
H I
LO
HOLD
Angle
rad deg
RECALL
RECALL
RECALL
RECALL
RECALL
ESC