131
12
SUPPLEMENTS
131
SU
PPLEMEN
T
S
Chapter
12
12-7.
Error/message list
■
Main numerical display section
-LoAd
Below the signal input range
-10V over (UTM
Ⅲ
/UTF)
-5V over (UTM
Ⅱ
)
The input electric signal greatly exceeds the signal input
range. This also occurs when an excessive load is applied to
the sensor or the input terminal is opened due to a break in
the cabling, etc.
LoAd
Beyond the signal input range
+10V over (UTM
Ⅲ
/UTF)
+5V over (UTM
Ⅱ
)
ALA
Beyond the alarm HI limit
This occurs when a larger load than expected is applied to
the sensor or calibration is performed below the
measurement range.
Check the calibration and measuring object.
-ALA
Below the alarm LO limit
oFL1
Below -99999
This occurs when a larger load than expected is applied to
the sensor or calibration is performed below the
measurement range.
Check the calibration and measuring object.
oFL2
Beyond 99999
cErr2
Beyond the zero calibration
range in the positive direction
The signal input range is greatly exceeded when zero
calibration is performed. Check for an excessive load
applied to the sensor and for breaks.
cErr3
Below the zero calibration
range in the negative direction
cErr4
Beyond the rated output value
in the positive direction/
negative direction.
Beyond the rated capacity
value.
The signal input range is greatly exceeded when calibration
is performed. In the case of equivalent input calibration,
check the rated output value, and in the case of actual load
calibration, check for an excessive load applied to the sensor
and for breaks.
Also, Check the setting of rated capacity value (display
value) sets it exceeding "±99999".
cErr5
Rated capacity value = 0
Check the setting of rated capacity value (display value) for
being "0" when calibration is performed.
cErr6
Rated output value 0.000V at
the time of equivalent input
calibration
Rated output value in the range
of -0.010 to +0.010V at the time
of actual load calibration
Check the rated output value for being 0.000V when
equivalent input calibration is performed, and the load for
being too light in weight when actual load calibration is
performed.
dZ Err
Digital Zero limit error
When the range of zero point correction (deviation from the
zero calibration registered point) by Digital Zero or Zero
Tracking exceeds the setting value of Digital Zero limit,
Digital Zero limit error occurs. Perform Digital Zero, etc., in
a proper range.
CALZE
Zero calibration in progress
These are not errors. They indicate that each calibration is in
execution.
CALSP
Actual load calibration
in progress
CALEq
Equivalent input calibration
in progress
485non
RS485 display mode
communication error
In RS-485 display mode, communication is being
established or format is not regularly received from UTM
Ⅲ
/
UTF. Please check if the connection is correct.
Содержание TM380
Страница 1: ...TM380 20MAY2021REV 1 02 TORQUE MONITOR OPERATION MANUAL ...
Страница 11: ...Contents X Contents X M E M O ...
Страница 147: ......