38
4
CALIBRATION PROCEDURE
38
CALIBRATION P
R
OCEDURE
Chapter
4
⑥
Zero clear (This step may be omitted if there is no change.)
Set the timing to reset the angle display to zero.
When reached to the specified number of rotations, it goes back to zero.
≪
Setting value
≫
- When the unit (angle) is [0: deg]
1 to 55 (It becomes 1 to 5 when the minimum scale is 0.1 to 0.5.)
- When the unit (angle) is [1: rad]
1 to 31
◇
Zero clear setting method
1.
Select setting mode 7.
Press
→
six
times.
2.
Set "Zero clear."
Press
six times.
⇒
"Zero clear" is displayed.
Press
, and the highest digit blinks.
Input the setting value with
.
Move to the lower digits with
. (Initial value: 55)
Validate it with
.
3.
Press
twice to go back to the indicated value display.
4-6.
Unit seal
For torque, affix a seal according to the unit at
which calibration has been performed.
Rotation speed does not matter either way.
Speed
rpm
Angle
rad deg
FNC
Torque
Speed
rpm
Nm
H I
LO
HOLD
Angle
rad deg
RECALL
RECALL
ESC
For torque
Unused
For rotation speed