101
10
OPTION INTERFACE
101
OP
TION IN
TERFA
C
E
Chapter
10
■
Output data select
《
Setting value
》
0: Torque (hold synchronized)
Torque synchronized with a hold is output.
1: Rotation speed
The rotation speed is output.
2: Angle
The angle is output.
3: Torque (hold not synchronized)
Torque in real time is output.
4: Zero fix
Output is fixed at the zero output (0V for voltage output, 4mA for
current output). After fixing, you will go to the output adjustment
mode, so that zero adjustment can be made.
See "
■
Adjustment by fixed output" on page 104.
5: Full scale fix
Output is fixed at the full scale (10V for voltage output, 20mA for
current output). After fixing, you will go to the output adjustment
mode, so that gain adjustment can be made.
See "
■
Adjustment by fixed output" on page 104.
◇
Output data select setting method
1.
Select setting mode 5.
Press
→
four times.
2.
Set "Output data select."
Press
once.
⇒
"Output data select" is
displayed.
Press
, and the highest digit blinks.
Press
three times to make the digit of the output
data select blink. (Initial value: 0)
Input the setting value with
.
Press
to validate it.
3.
Press
twice to go back to the indicated
value display.
CAUTION
●
Cable can be from 24 to 14AWG (0.2 to 2.5mm
2
).
●
It is not necessary to solder the cable wires or to fix a solderless terminal.
●
If several cables to be inserted to the same hole, twist those cable wires together
and insert.
●
Connect the cable with the power supply of the main body turned off without fail.
Speed
rpm
Angle
rad deg
FNC
《
Setting value
》
Undefined
0 0 0 0
Undefined
Undefined
Output data select
0: Torque
(hold synchronized)
1: Rotation speed
2: Angle
3: Torque
(hold not synchronized)
4: Zero fix
5: Full scale fix
Torque
Speed
rpm
Nm
H I
LO
HOLD
Angle
rad deg
RECALL
RECALL
ESC