90
10
OPTION INTERFACE
90
OPT
ION INT
E
RFA
C
E
Chapter
10
2. Send/receive [setting value writing/reading]
* Command definition
Setting value writing
Function
Header
Protocol
Main
Sub
Various settings
W
-
Setting value reading
Function
Header
Protocol
Main
Sub
Various settings
W
-
W
CR
Host
TM380
Setting value (sign, 5-digit, no decimal place)
Command
(For a setting value with a decimal place, simply write
the numerical value excluding the decimal place.)
W
CR
* When the number of digits of the setting value is smaller than five,
Example) 3 digits
* *
+
0
0
1
2
3
W
CR
* When the setting value is unsigned, put a zero in the sign digit.
* *
0
0
0
0
0
3
put a zero in the higher digit(s).
W
Setting value
Command
(A setting value with a decimal place results
in the numerical value excluding the decimal place.)
Delimiter
W
CR
Host
TM380
Command
(sign, 5-digit, no decimal place)
* The mode number of the setting value of HI/LO limit is "0."
* For setting mode numbers and setting item numbers,
see "12-1.The list of initial setting value" on page 118.
*
*
Setting value item number
Setting mode number (F*)
- "Initialize" and "password" cannot be either read or written.
- When the protect of the setting value you are going to write is ON, it cannot be written.
Turn OFF the protect.
- If an output value is written in command 42 (zero calibration), simulated zero calibration
is executed with the written value. Also, if an output value is written in command 43
(equivalent input calibration), equivalent input calibration is executed.
Point