12-3.
Setting protect
Each category of setting values can be locked to prevent rewriting.
For each category, see the setting value list.
《
Setting value
》
0: OFF
Enables rewriting of setting values.
1: ON
Disables rewriting of setting values.
■
Setting protect
setting method
1.
Select setting mode 9.
Press
→
seven times.
2.
Set "setting protect."
Press
once.
⇒
"Setting protect" is displayed.
Press
, and the highest digit blinks.
Input the setting value of "Work protect" with
.
Move to the lower digit with
. (Initial value: 0)
Input the setting values of "System protect" and
"Calibration protect" in a like manner.
Press
to validate it.
3.
Press
twice to go back to the indicated value
display.
Speed
rpm
Angle
rad deg
FNC
Torque
Speed
rpm
Nm
0
0
0
Calibration protect
0: OFF 1: ON
System protect
0: OFF 1: ON
Work protect
0: OFF 1: ON
《
Setting value
》
H I
LO
HOLD
Angle
rad deg
RECALL
RECALL
ESC
For the setting items protected by the setting protect, see "12-1.The list of initial
setting value" on page 118.
Point