4.Methods of Connection
22
4-1-1-2. Strain Gauge Sensor Connection
Before connecting a sensor, set the excitation voltage, turn off the power, and perform
the following connection. (See "5-6. Excitation Voltage" P34.)
・
4-wire sensor
・
6-wire sensor
For connecting a 6-wire strain gauge type sensor, short-cEXC and +S, and -EXC
and -S.
・
For connecting a sensor, be sure to use shielded lines, and separate
wiring from lines with heavy noise (such as wiring of power equipment
and wiring of digital equipment) and AC lines.
・
The analog GND in the F395 is grounded to F.G. Also, when voltage/
current input is specified, the -SIG terminal is at about the same level
as the analog GND. (See "20. Block Diagram" P152.) Carry out wiring
so that unnecessary current will not flow.
Request
+
EXC
-
EXC
+
SIG
-
SIG
F.G
+
IN
-
OUT
-
IN
+
OUT
F395
+
EXC
-
EXC
+
SIG
-
SIG
F.G
+
IN
-
OUT
-
IN
+
OUT
+
S
-
S
F395
Содержание f395
Страница 1: ...Operation Manual F395 DYNAMIC FORCE PROCESSOR 10 Apr 2012 Rev 1 12...
Страница 167: ...21 Outer Dimensions 153 21 Outer Dimensions Unit mm 3 3 5 5 Rear Front Side Panel cut dimensions...
Страница 180: ...Unipulse Corporation 9 11 Nihonbashi Hisamatsucho Chuo ku Tokyo 103 0005 Tel 81 3 3639 6120 Fax 81 3 3639 6130...