4.Methods of Connection
27
・
CODE1 ~ 16:
Specifies operation ch for Hold mode, High/Low Limit, Comparison Waveform, etc.
(See "10-2. Setting and Operating Method" P47, "11-2. Setting and Operating
Method" P64, "12-2. Setting and Operating Method" P65, and "13-2. Setting and
Operating Method" P68.)
・
STROBE:
Changing of operation ch by changing CODE 1 ~ 16 is prohibited when this signal is
ON. (Previous operation ch is maintained.) Use to prevent chattering when changing
operation ch.
・
Select output:
Output pin 44 ~ 46 (LO/OK/HI displacement 2) operate when this signal is ON.
- Pin 44/45 "Status" output
(
For logic, see "4-3-4. Equivalent Circuit (Output)" P30. )
- Pin 46 "Level Edge" output (Present when this signal is ON. Use for level
edge clear, etc.)
When this signal is OFF, output pin 44 ~ 46 result in the operation of LO/OK/HI
displacement 2.
Judgment output signals
The results of High/Low Limit Judgment in each Operation Mode are output.
・
LO1, OK1, HI1
:
1) When the Multi-Hold / Hysteresis mode is used.
Outputs the results of judgment of the load with respect to the
high limit / low limit.
2)When the Waveform Comparison / Waveform and Displace-
ment Comparison mode is used.
Outputs the results of judgment of the load with respect to the
high and low limit waveforms.
3) When the Hysteresis 2 mode is used.
Outputs the results of judgment of the load with respect to the
high limit / low limit, and the results of judgment of the
difference between go and return with respect to the hi-hi
limit / lo-lo limit.
Status
Pin 44
Pin 45
WAITS / STOP
0
0
WAIT2
1
0
WAITL
0
1
SAMP / COPY
1
1
A level edge is generated when the input signal (indicated value)
crosses over the Waveform Start Level after the START signal is input.
It is cleared by starting sampling or inputting the STOP signal. (See "9.
Method of Starting Measurement" P43.)
Содержание f395
Страница 1: ...Operation Manual F395 DYNAMIC FORCE PROCESSOR 10 Apr 2012 Rev 1 12...
Страница 167: ...21 Outer Dimensions 153 21 Outer Dimensions Unit mm 3 3 5 5 Rear Front Side Panel cut dimensions...
Страница 180: ...Unipulse Corporation 9 11 Nihonbashi Hisamatsucho Chuo ku Tokyo 103 0005 Tel 81 3 3639 6120 Fax 81 3 3639 6130...