13.Waveform Comparison / Waveform and Displacement Comparison Mode
70
2. Set 100Hz under "Sampling Rate". However, it is normally 2kHz in the Displacement
mode. For use in the Displacement mode, operations in "5." ~"9." are unnecessary,
therefore perform operation in "10." after "3." and "4.".
3. Set "Waveform Start Level" and "Waveform Termination Level" at the desired ch (0 ~
15).
Waveform Start Level: Normally set a value close to zero (approx. the increment
×
10). Set a slightly larger value if noise is large.
4. Set desired channel(s) with external channel selection signals.
5. Exit the setting menu, go to the main screen, input the START signal externally, and
then input a test waveform.
MODE
→
Operation Mode
→
Sample Rate
MODE
→
Code Set
→
Basic: Wave Start Level / Wave End Level
10
COM
11
CODE1
12
CODE2
13
CODE4
14
CODE8
36
CODE16
Open CODE16.
(Do not use.)
Request
Example: 5ch
Control connector
A waveform is input in synchronization with the
"Waveform Drawing" function.
For details, see "9. Method of Starting Measurement" P43.