16.Code Setting
116
16-2-11. Minimum Peak and Valley Difference /
Peak and Valley Detection Rate / Peak and Valley Ordinal
This is displayed only when Relative Minimum Value / Relative Maximum Value Hold
is selected in Hold Setting to set the relative maximum and relative minimum value
detection parameters. However, most waveforms can normally be handled at factory-
shipped settings. Set referring to the following principles of operation only in cases
where holding is unsuccessful or finer adjustments are needed. Relative maximum and
relative minimum values are detected based on the following principles.
Minimum Peak and Valley Difference
Setting range: 1 ~ 9999 (Not including the decimal point.)
Peak and Valley Detection Rate
Setting item: 25% / 50% / 75% / 125% / 150% / 200% / 300% / 400%
Peak and Valley Ordinal
Setting item: 1 ~ 9
MODE
→
Code Set
→
Basic: Hold Set
→
P&V Difference
MODE
→
Code Set
→
Basic: Hold Set
→
P&V Judg. Rate
MODE
→
Code Set
→
Basic: Hold Set
→
P&V Ordinal