4.Methods of Connection
23
4-1-1-3. Voltage / Current Type Sensor Connection
Contact type, overcurrent type, laser type, and other voltage / current type sensors of up
to
±
5V /
±
20mA can be connected. (The analog input option is a factory-shipped
setting.)
4-1-1-4. Voltage Output (V-OUT) Connection
The relationship between sensor type and output voltage is as follows:
Strain gauge
1mV/V ............ Approx. 2.0V
Voltage
5V ............. Approx. 5.6V
Current
20mA ............. Approx. 5.2V
4-1-2. SI/F Connection
This is a 2-wire serial interface to connect a UNIPULSE-manufactured printer, external
display, etc.
Up to three nonpolarized external devices can be connected. Use parallel two-core
cables, cabtyre cables, and the like for wiring.
+
SIG
-
SIG
Voltage / current
+
-
F395
Output
type sensor
Within
±
5V/
±
20mA
V-OUT
GND
External equipment
Load resistance
5k
Ω
or more
+
-
F395
No displacement sensor (for X-axis measurement) can be
connected when the PUI option is mounted.
Converter
Display
Printer
SI/F
F395
<Example of connection>
Содержание f395
Страница 1: ...Operation Manual F395 DYNAMIC FORCE PROCESSOR 10 Apr 2012 Rev 1 12...
Страница 167: ...21 Outer Dimensions 153 21 Outer Dimensions Unit mm 3 3 5 5 Rear Front Side Panel cut dimensions...
Страница 180: ...Unipulse Corporation 9 11 Nihonbashi Hisamatsucho Chuo ku Tokyo 103 0005 Tel 81 3 3639 6120 Fax 81 3 3639 6130...