INTRODUCTION
8
USER MANUAL — 200-9011-ENG R1
Environmental conditions
Use
ambient temperature
relative humidity
atmospheric pressure
50 °F to 104 °F
5 to 95% without
condensation
500 to 1060 hPa
10 °C to 40 °C
5 to 95% without
condensation
500 to 1060 hPa
Transport and storage
ambient temperature
relative humidity
atmospheric pressure
-40 °F to 158 °F
10% to 100%
500 hPa to 1060 hPa
-40 °C to 70 °C
10% to 100%
500 hPa to 1060 hPa
Maximum altitude
6,562 ft
2000 m
Electrical requirements
Electrical and mechanical requirements
Compliance with standards: IEC60601-1, CAN/CSA C22.2
60601-1, UL60601-1, IEC60601-1-2, IEC60601-2-52,
CAN/CSA C22.2 60601-2-52, IEC60601-1-6:2010,
CAN/CSA C22.2 60601-1-6:11
Degree of protection
IPX4
Degree of protection against electrical shocks
Class I, Type B
Scale accuracy
1 % with a difference of maximum 3 lb / 1.4 kg
Voltage
120 V AC, 60 Hz / 220-240 V AC, 50-60 Hz
Amperage (includes power outlet and powered system)
120V AC: Bed 4.25A / Auxiliary outlets (optional) 5A /
Total = 9.25A
220-240V AC: Bed 2.15A / Auxiliary outlet (optional) 2A /
Total = 4.15A
Current leakage
<300 micro-amperes
Battery type
sealed rechargeable lead acid battery 1.2 Ah, 2 x 12 V
Actuator duty cycle
1 minute On, 9 minutes Off
Auxiliary power outlets
120V AC / 5A
230V AC / 2A
USB type A charger
120V bed: 5V DC / 3.1A
220-240V bed: 5V DC / 0.75A