OPERATING INSTRUCTIONS
26
USER MANUAL — 200-9011-ENG R1
Power cord support brackets
Support brackets are available at the head end of bed (illustrated) to store the power
cord during transportation or when not in use.
CPR release
CPR release handles are located on each side of the bed frame, under the backrest section. Usage shall be
limited to emergency situations only. An over usage could lead to mechanical degradation.
Operation:
1.
To activate, pull and hold the CPR release handle to flatten all sections of patient sleep surface.
2.
Release handle once surface has reached flat position. If released before surface is completely flatten, all
sections will stop retracting.
WARNING
• CPR release option usage shall be restricted to emergencies. To avoid injuries or product damages, ensure
total clearance around product prior to usage.
ATTENTION
• Usage shall be limited to emergency situations only. An over usage could lead to mechanical degradation.
•
If bed operates on battery, the thigh section will not reach flat postion.
IMPORTANT
Thigh section will not reach flat position if master lockout of thigh section or, a movement function of any control
is activated during motion.
Nurse call connector
If the bed is equipped with the optional nurse call system, the connector selected when ordered will be located at
the head end of the bed, on the frame. The connector is only designed for nurse call function and for room lights
control. Communication will be sent through system for a Nurse call initiated by the patient and/or for a bed exit
control alarm. Ensure the plug is properly inserted into connector.
!
!