OPERATING INSTRUCTIONS
38
USER MANUAL — 200-9011-ENG R1
• Auto Arm in Arm Detection
to arm
automatically the bed exit detection when
patient ingresses in bed. If user selected the
standard interface, the Auto Arm button is not
displayed on the home screen.
Configuration steps:
1.
From the Home screen, press the Arm
Detection button.
2.
Select appropriate answer to the prompt
question message box in order to arm or
disarm the Auto Arm detection option.
3.
When selecting the Auto Arm option, AUTO will appear on the upper right corner of the Arm Detection
button on the screen.
4.
When Auto Arm is activated and patient is detected in bed, a message box will appear to indicate
that the patient has 60 seconds to settle in bed before the bed exit is armed automatically. If patient
is not settled after 60 seconds or if patient gets out of bed before 60 sec. is over, bed exit detection
system will fall in alarm.
Home
November 1st 12:00
Arm
Detection
Detection 3
Fall Risk
Mgmt
Pref.
Auto
Arm
0
º
30’’
0 º
AUTO
No
Yes
Arm Bed Exit ?
Auto
Arm