GETTING STARTED
20
USER MANUAL — 200-9011-ENG R1
17. If applicable, test the room lights functions for proper operation.
18. If applicable, verify following options for proper operation: lateral bed extension, auxiliary power outlets and
USB port, nightlights, etc.
19.
Install mattress with proper specifications.
20. If bed is equipped with bed exit detection, perform a zero procedure on the bed after all equipment has been
installed.
Zeroing procedures
ATTENTION
• Patient must be out of bed.
Zeroing procedure for a bed with scale
Cancellation of zeroing process can be done at any time by pressing the “Cancel” button.
Procedure:
1.
Bed is electrically powered up. Ensure sleep surface is completely flat and put the bed elevation system at
intended height of use. Apply the brakes.
2. From the display screen on the footboard, enter the Scale menu.
3. Press the “zero” button.
4. Password may be required, enter password.
5. Follow the instructions; patient out of bed, install all required equipments, accessories and mattress.
6. Press ZERO button on display screen again.
Do not touch bed while this is in progress
.
7.
Zero is successful if a confirmation message appears and the display screen returns to “Scale” menu.
8. If failure occurs, a message appears to indicate that weight is unstable. Display screen will stay in Zero menu.
9. Ensure there is no interference and start over the procedure.
!