Care Pause
When the bed exit detection is armed, press the Disarm Detection button to apply a Care Pause and give the
caregiver 5 minutes to provide any care to the patient without disactivating the bed exit system and without
generating a bed exit detection audible sign. The detection button on the screen will turn orange and the time left
for the pause will appear. If applicable, the sideview lights bumpers will turn solid orange for the duration of the
pause.
Select:
• Care Pause
to apply a 5 minutes pause to the
armed bed exit detection.
• Cancel
to return to the Home screen.
Anytime during the pause, the time can be resumed
to 5 minutes if needed. When 30 seconds remain,
an audible sound will be heard and the time left will
be displayed and flashing in the screen. When only
10 seconds remain, another audible sound will be
heard.
Select:
• Deactivate Pause
to deactivate the pause
completely.
• Resume to 5 min.
to reset the time to a total
of 5 minutes.
• Close
to keep the remaining time the same
and to return to the Home Screen.
When Auto Arm is activated and the pause finishes,
the system will start the auto detection. When
detected, the patient has 60 seconds to settle in the
bed before the bed exit is armed automatically.
When auto-arm is activated and bed is in alarm while
patient is still in bed, press the Disarm Detection
button to stop the alarm. The system will automatically
apply a one-minute pause to the auto-arm to allow
the caregiver to attend to the patient. Press on the
Detection button to apply a 5-minute care pause if
needed. If time runs out after one minute, the bed
automatically returns to the auto-arm mode. Refer to
the
Home screen with integrated scale
section (page 37).
November 1st
12:00
Arm
Detection
Detection 3
Scale
Pref.
Fall Risk
Mgmt
0 º
30
"
0 º
Auto-arm activated
and patient in bed.
Cancel
Care
Pause
November 1st
12:00
Arm
Detection
Detection 3
Scale
Pref.
Fall Risk
Mgmt
0 º
30
"
0 º
Care pause activated
Deactivate
Pause
Close
Resume to 5 min.
November 1st
12:00
Pause
2:30
Scale
Pref.
Patient
Risk Mgmt
0 º
30
"
0 º
Detection 2
AUTO