5-6
5.5 Replacing consumables
5.5.1 Fuses
Disconnect the power cord from the outlet when
replacing fuses. Otherwise, you may get an electric
shock, resulting in death or serious injuries.
Use the fuse (250 VAC, T2.5AL) specified on the label
on the side of the fuse holder. Always replace two
fuses together. Using another type of fuse may result
in failure or fire.
When the instrument does not work correctly after
fuses have been replaced, there may be other causes
of the problem. Turn off the instrument immediately
and contact our local distributor.
1) Turn power off and disconnect the power plug from the
outlet.
(2) Disconnect the connector of the power cord from the
outlet.
3) Insert a coin or similar in the slot on the fuse case (1)
located on the side of the main unit. Turn the screw
counterclockwise to remove it.
(4) Replace the blown fuse with a new one.
(5) Reverse the procedure above to install the fuse holder.
5.5.2 Chin rest paper
1) Remove the two chin rest paper pins.
2) Place new chin rest paper on the chin rest and secure
the paper with the paper pins.
(1)
(Fig. 1)
(Fig. 1)