3-118
●
Unit
Select the standard unit for displayed data: D
(diopter) or mm (millimeter).
●
Correction Mires
No data is displayed on the RING map if mires
cannot be detected on the cornea. However, when
this option is set to ON, the instrument places
corrected mires as blue dots in the area not shown
on the map.
●
Transmissivity
Select the default transmissivity of the map.
●
Map Interpolation
Select whether to display or hide (ON/OFF)
interpolated sections of the map.
When “ON” is selected, sections of the map that
could not actually be analyzed are interpolated and
displayed.
●
Slit Offsets
Select whether to display or hide (ON/OFF) the
offset of the slit image.
● Elevation BFS
Define settings for calculating the elevation BFS.
Select “Float,” “Axial,” or “Apex” for BFS Alignment.
Set BFS Diameter within the range from φ0.1 mm to
φ13.0 mm.
●
Display Index
Set the default quantitative index to be displayed on
the Single Map or 6 Maps screen.
●
Map display setting