3-62
re-calculation is required.
If the curvature at the corneal center is even, the EDP
and EDD are essentially “0.”
d) Keratoconus Screening
When diagnosing keratoconus in practice, check
thinning of cornea, Fleischer ring, Vogt's striae
(Keratoconus line), etc. as per normal examination.
The Keratoconus Screening System is an effective
program for diagnosis and observation of keratoconus.
This program is designed to analyze whether there is a
pattern unique to keratoconus on the color map using two
approaches and to evaluate the corneal shape.
To be displayed on the Keratoconus Screening tab, the
following message appears: This program does not screen
all abnormal conditions of the cornea. This program is
designed to detect the keratoconus by distinguishing the
keratoconus pattern shown on the corneal shape from the
normal cornea or other corneal conditions.
Click the “OK” button to display the map and index.
(Fig. 1)