i Important safety information
Do not install this instrument in a location where explosives or inflammable
substances are used or stored. Fires or explosion may occur.
Do not remove the cover of the instrument. Otherwise, you may be directly exposed
to high voltage sections.
Do not disassemble or modify the instrument. Otherwise, you may be directly
exposed to high voltage sections.
The personal computer specified in this manual and an Isolation transformer that
isolates the power source for the personal computer from the commercial power
source are required to use this instrument. (No Isolation transformer is required
when using a personal computer specifically designed for medical equipment.) Be
sure to observe the following precautions in relation to your personal computer
and Isolation transformer in order to safely use them as part of a medical
equipment system. Refer to “8. Specifications” for the performance requirements
of the personal computer and Isolation transformer. Contact our local distributor
or Tomey Corporation if you have any questions.
Use a personal computer conforming to IEC60950-1 and an Isolation transformer
conforming to IEC60601-1. Otherwise, they may not function safely as part of a
medical equipment system.
Connect the power cord of the personal computer to the Isolation transformer, not
directly to a commercial power source. Otherwise, they may not function safely as
part of a medical equipment system.
Disconnect the power cord from all the connected devices before installing and/or
servicing this instrument. Otherwise, you may get an electric shock.
Do not place the Isolation transformer directly on the floor when in use.
Do not connect a plug-in socket or extension cord to the Isolation transformer.
Only use the specified terminal for connection to the main unit of this instrument.
Using any other type of terminal may result in failure of the instrument.
Do not connect electric appliances not provided in the system package to the
Isolation transformer. Otherwise, failure may occur to the Isolation transformer or
electric appliances.
Содержание TMS-5
Страница 2: ......
Страница 6: ...i 4 This page is intentionally blank...
Страница 24: ...1 12 This page is intentionally blank...
Страница 97: ...3 47 Fig 1 Fig 2...
Страница 127: ...3 77 RING TOPO Viewer Ks Kf MinK AveK Es Em CYL SRI PVA SAI...
Страница 186: ...4 6 This page is intentionally blank...
Страница 202: ...7 2 This page is intentionally blank...
Страница 214: ......
Страница 216: ......