3-16
3.4.4 Capturing in slit mode
Faulty fixing of the sight, eyelid drooping, trichiasis,
corneal diseases, etc. may affect the measurement
result.
When the measurement image is unclear, select or
measure another image.
a) 3D scan
3D scan data is obtained by capturing images of a cross
section of the anterior chamber in series, while turning the
capturing position.
1) Select the slit mode. (Refer to “3.3.3 Selecting the
capture mode.”)
2) Use the joystick to position the patient's eye on the
screen.
3) Align the reflection (1) at the corneal center with the
center of the center ring (2).
4) Tilt the joystick back and forth so that the focus
indicator (3) becomes smaller. When the focus
indicator is shown horizontally, the head is too far from
the patient's eye; when the focus indicator is shown
vertically, the head is too close to the patient’s eye.
5) When Auto Shot (4) is ON, capturing automatically
starts.
When Auto Shot (4) is OFF, press the joystick button (5)
to start capturing.
(Fig. 1)
(4)
(2)
(1)
(3)
(Fig. 2)
(5)
(Fig. 3)