7. RESSETING COUNTER IN ALARM SYSTEM / ONLY MODEL AP-POOL-X1/
COUNTER IN THE ALARM SYSTEM SHOULD BE RESSETED EACH TIME THE UV
LAMP IS REPLACED
a) Disconnect from power supply /neutral and life/.
b) Replace UV Lamp
– see p.
9.
INSTRUCTION FOR REPLACEMENT OF UV LAMP
c) Connect device to Power supply (power switch should be turned off)
d) Hold the right button under display.
0
1
NACISNĄĆ I PRZYTRZYMAĆ
OK
CZEKAJ 10 SEKUND
e) Turn on the power switch while still holding the right button for more than 10 seconds.
f) Short sound signal will confirm resetting the work time counter.
8. LEGEND FOR INDICATOR LIGHT ON ENCLOSURE
Lighted color indicator lights inform the operator of the status of the sterilizer. Certain
conditions can be identified from the configuration of lights. For these corrective actions are
listed below:
Alarm
Red Light
Power Supply
Green Light
Power
Green Light
Problem
Action
OFF
ON
ON
No fault
-
OFF
BLINKING
ON
No fault
– heating the
UV lamps
-
OFF
OFF
OFF
No Power
Check power
source
ON
OFF
ON
1.Burned out or faulty
UV lamp
2.Faulty starter
1.Replace UV
lamp
2.Replace starter
OFF
OFF
ON
Alarm System Fault
Contact Dealer.
Replace alarm
system.