CLD101x
5.3 Troubleshooting
Strange display messages
The CLD101x shows unexpected display values, as for example:
·
The laser diode might not be installed correctly and/or the wrong pin type is selected.
ü
Check the
to determine type1 or type2 standard pin-out.
ü
correctly.
Cannot switch on the TEC
·
NTC resistance measurement resulted in invalid values (too high or too low values)
·
Temperature reading shows "FAILURE" (NTC "OPEN" or "SHORT")
·
An error message "TEC output voltage protection was tripped - ..." appears, TEC does not
switch on: The laser current circuit seems to be interrupted.
ü
Check the laser pin-out - only
type1 and type2 can be used
ü
Check if the laser is installed
(pin type 1 or 2, pin 1 at correct location)
ü
The laser diode may be defective
Cannot switch on the laser
·
NTC resistance measurement resulted in invalid values (too high or too low). Temperature
reading shows "FAILURE" (NTC "OPEN" or "SHORT")
ü
Check the laser pin-out - only
type1 and type2 can be used
ü
Check if the laser is installed
(pin type 1 or 2, pin 1 at correct location)
ü
The laser diode may be defective
·
When the laser is switched on, the display shows LASER is PAUSED
ü
feature "Pause Laser" was activated due to laser temperature
out of range. The display also shows
and
75
10
75
10
75
10
22
Содержание CLD101 Series
Страница 1: ...Compact Laser Diode Controller CLD101x Operation Manual 2018 ...
Страница 2: ...Version Date 1 7 09 Jul 2018 Copyright 2018 Thorlabs ...
Страница 76: ... 2018 Thorlabs 74 CLD101x 6 2 Dimensions CLD101x ...
Страница 84: ... 2018 Thorlabs 82 CLD101x 6 5 Certifications and Compliances ...
Страница 90: ......
Страница 91: ......
Страница 92: ...www thorlabs com ...