4 Write Your Own Application
57
Bit #
Description
13
Socket connection failure (Socket missing or unknown socket detected)
12
reserved, read as 0
11
TEC output compliance voltage reached
10
Temperature sensor failure.
9
Temperature window failure.
8
Temperature protection is active.
7
TEC current limit reached
6
reserved, read as 0
5
reserved, read as 0
4
reserved, read as 0
3
LD current limit reached
2
LD interlock is active
1
LD output compliance voltage reached
0
Keylock protection is active
4.3.4 Error Reporting
The device stores errors in a queue containing up to 10 entries. The error queue may be read out
by the
‘SYSTem:ERRor[:NEXT]?’
command. The following table lists all error numbers and the
according descriptive messages. Note: negative numbers are defined by SCPI while positive error
numbers are device dependent.
Error
Description
0
No error
1
The error couldn't be specified more precisely
2
Floating point domain error
3
Device temperature too high
4
General GUI error
5
Authentication required for operation
6
Authentication process failed
7
Operation is not allowed in service mode
8
Operation is allowed in service mode only
9
A measurement is currently in process
20
Operation not allowed while LD output is on
21
Wrong operating mode for this operation
22
INTERLOCK circuit is open
23
KEYLOCK is active
24
Operation not allowed because of a 'OPEN CIRCUIT' condition
Содержание CLD101 Series
Страница 1: ...Compact Laser Diode Controller CLD101x Operation Manual 2018 ...
Страница 2: ...Version Date 1 7 09 Jul 2018 Copyright 2018 Thorlabs ...
Страница 76: ... 2018 Thorlabs 74 CLD101x 6 2 Dimensions CLD101x ...
Страница 84: ... 2018 Thorlabs 82 CLD101x 6 5 Certifications and Compliances ...
Страница 90: ......
Страница 91: ......
Страница 92: ...www thorlabs com ...