95
Thermo Fisher Scientific
User Guide STP420D/STP420ES
troubleshooting and Errors
Fill or Drain Problems
8. Touch
OK
again to confirm that the Xylene cleaner bottle is in
position X.
9. Press
OK
again to continue the station cleaning program.
Station cleaning proceeds with a series of messages:
• After the first step of the protocol
:
“Please put the Alcohol cleaning bottle in position X, the Xylene
cleaning bottle in position 11 and the cleaned bottle in position 12.”
Follow the instructions.
Note!
Position 11 is the second from the right in the bottom row, right next to
position 12
.
•
“Please confirm that alcohol cleaner bottle is in position X”.
Touch
OK
to confirm this. The cleaning program will resume. No
further input required before the program is finished.
• “Emptying: cleaning program is finished, do you want to open the lid?”
Click
OK
to accept
•
“Rotation program: Lid is unlocked and can be opened”
.
Touch
OK
to acknowledge this.
•
“Station cleaning: cleaning for station X is deleted.”.
Touch
OK
to acknowledge this.
•
“Please put the cleaned bottle and the Alcohol cleaning bottle X back in
their original positions.”
Touch
OK
to confirm you have done this.
The station cleaning display appears on the screen, then the Standby
screen returns.
Note!
If you want to perform a station cleaning session without getting an error
message, you can run the cleaning session as long as no protocol is running.
To do this press the following keys:
“Settings”, “Cleaning”, “Station Cleaning” and choose the station and the
desired chamber.
Note!
When performing the cleaning of must a station 17 External Tank, the
external tank must be used instead of the Xylene cleaning bottle. In this case
the external tank must hold at least 6 liters of Xylene.