76
Thermo Fisher Scientific
User Guide STP420D/STP420ES
6. Reinsert the filled bottle into its correct location. Make sure you
hear the ‘click’ that confirms it is correctly placed.
7. Inform the system that you have replaced a reagent:
a. Press
Reagents
.
b. Press the appropriate reagent container icon on the screen. The
icon is marked with a frame, for example:
c. Press
Clear Counter
.
Note!
Replacing reagents: You can only replace reagents automatically if you have
‘supervisor’ access.
Note!
Draining and refilling reagents: Reagent stations drain and refill through
the rotation chamber. Before the automated reagent replacement starts, you
must carry out a clean cycle.
1. Place a suitable waste container in front of the instrument.
Note!
Automatic replacement: Make sure the waste container is large enough for
the quantity being drained, and is suitable for the reagent.
2. Attach the drain hose to the gray connector socket marked ‘Ext
Tank’ on the front of the equipment:
Formalin 1
3-6-3 automatically
replacing reagents
Drain the reagent
routine operations
Cleaning Baskets and Replacing Reagents
Figure 34.
Frame marking a selected reagent.
Figure 35.
Outlet for draining the paraffin tank.