109
Thermo Fisher Scientific
User Guide STP420D/STP420ES
transportation and Final Shutdown
Transportation: Preparing for Shipment
4. Empty the reagents.
5. Empty the paraffin.
6. Empty the wash bottle (see Section 3-3-2 Emptying water from the
wash bottle).
7. Turn the instrument off, as described in Section 3-1-2 Turning the
instrument off.
8. Unplug the unit.
9. Dispose of all liquids appropriately, according to the safety data
sheets and any applicable laboratory regulations.
10. Install transport handles as described in Section 2-3 Transporting
the STP420 to Its Site.
11. Pack the instrument
in the original packaging
, as this gives you the
best chance of transporting the instrument without damage.
Note!
Original packaging: If the original packaging is no longer available, contact
your local Thermo Fisher Scientific representative.
r
eturninG
the
instrument
If you return the unit, make sure it is clean and safe.