Notify an authorized service technician if a problem or sus-
pected problem occurs. The instrument must be serviced prop-
erly by a fully authorized representative to ensure it continues
to comply with applicable safety precautions.
f
lammaBle
enVironments
Never operate the instrument in the presence of flammable
gases. After filling up the reagents no ignition sources must be
near the device or in its vicinity.
m
alfunCtion
hazarD
To avoid severe software and/or mechanical failure, do not
turn off the instrument without following the instructions
in Section 3-1, Turning the Instrument On and Off. Any dam-
age caused by not turning the instrument off properly is ex-
cluded from warranty, and any repair and/or service call will
be charged.
B
ioloGiCal
hazarD
Specimens used during the intended operation of the instru-
ment might be infectious. You are strongly advised to observe
the general laboratory regulations concerning protection
against danger of infection. You can find information on
decontamination media, and their use, dilution and effective
range of application, in the Laboratory Biosafety Manual: 1984
of the World Health Organisation.
u
nhealthy
or
irritatinG
suBstanCes
Small amounts of irritating aerosols are possible during the
intended use of the instrument. To avoid risk to your health
while you are in direct contact with the reagents or their resi-
dues, make sure there is sufficient ventilation, and always wear
appropriate safety clothing (for example, protective glasses,
laboratory gloves and so on).
r
eaGents
You must operate the stp420 only with the reagents listed in
Appendix D, Approved Reagents.
C
leaninG
reaGent
–
paraffin
BloCkaGes
The cleaning reagent must only be Xylene or Clear-Rite3. Oth-
er Xylene substitutes (such as Toluol or others) might lead to
paraffin blockages and must not be used. Any damage caused by
using any cleaning reagent other than Xylene is excluded from
warranty, and any repair and/or service call will be charged.