User Guide STP420D/STP420ES
3-5-2 adding more
specimens while a
protocol is running
routine operations
Starting a Protocol
The addition of more samples during a protocol run extends the process
by the same amount of time that the interruption lasts.
1. From the main menu, select
Rotation
or
Biopsy
, as appropriate.
2. Select
Open
.
3. Select
Yes
in response to “Do you really want to open the chamber?”
f
lammaBle
reaGents
If the reagent is marked as flammable in reagent management, the
instrument drains the reagent before the lid is released, when you try to
open the chamber.
4. Press
OK
in response to “Rotation/biopsy chamber lid is unlocked
and can be opened”.
5. Open the lid, add the cassettes, and close the lid.
6. Select
Continue
.
7. Select
Yes
in response to “Do you want to continue the program?”
The chamber fills with reagent, as necessary, and the process continues.
Note!
The “Open User” or “Open supervisor” applies to all protocols in each
chamber (see section 3-7-3 Creating a new protocol).
Note!
After you have pressed Continue, a start-up procedure runs for a short time.
you can not finish the protocol during that.
Note!
If the protocol does not continue after 2 minutes, a message appears on the
screen and an alarm will sound.