![Thermo Scientific STP420D Скачать руководство пользователя страница 119](http://html.mh-extra.com/html/thermo-scientific/stp420d/stp420d_instruction-manual_1104825119.webp)
119
Thermo Fisher Scientific
User Guide STP420D/STP420ES
appendix F
Preferred Protocols and Best Practices
Program step
time (mins)
Pressure/
vacuum (P/V)
temp (0°C) Drain time
(secs)
time to fill and
empty (mins)
Standard day:
1
Formalin
30
P/V
0
60
4
2
Formalin
30
P/V
0
60
4
3
Alc 80%
40
P/V
0
20
4
4
Alc 95%
20
P/V
0
20
4
5
Alc 95%
20
P/V
0
20
4
6
Alc 100%
40
P/V
0
20
4
7
Alc 100%
65
P/V
0
60
4
8
Xylene
30
P/V
0
20
4
9
Xylene
45
P/V
0
60
4
10
Paraffin
30
P/V
62
20
4
11
Paraffin
30
P/V
62
20
4
Total time including time to fill
and empty (hr:min)
08:19
reagent
time (Minutes)
Xylene cleaner
10
Xylene cleaner
10
Alcohol cleaner
5
Cleaning cycle “Short Cleaning”
(both chambers)
reagent
time (Minutes)
Xylene cleaner
10
Xylene cleaner
10
Xylene cleaner
10
Alcohol cleaner
5
Cleaning cycle “Standard
Cleaning” (both chambers)
reagent
time (Minutes)
Xylene cleaner
1
Alcohol cleaner
1
Cleaning cycle “Cleaning reagent
station” (station cleaning, both
chambers)