![Thermo Scientific STP420D Скачать руководство пользователя страница 117](http://html.mh-extra.com/html/thermo-scientific/stp420d/stp420d_instruction-manual_1104825117.webp)
117
Thermo Fisher Scientific
User Guide STP420D/STP420ES
appendix E
Technical Data
Software
integrated PC and LinUX
operating system
Rotary motor R7
P24.7
F9
6.3 A
Power board A3:
Instrument fan
M1/M2
F1
0.2 AT
Rotary chamber fill/Heating of the
medium pipe
R21
F2
1.6 A
Outlet paraffin heater
O12
F3
1.6 A
Rotary chamber heater, lower front
R12
F4
4 A
Rotary chamber heater, lower
R15
F5
12.5 A
Rotary chamber heater, upper
R28.1
F6
10 AT
Biopsy chamber heater
B12/B13
F7
12.5 A
Paraffin container heater 1
O6.1
F8
6.3 A
Paraffin container heater 2
O6.2
F9
6.3 A
Paraffin container heater 3
O6.3
F10
6.3 A
Paraffin container heater 4
O6.4
F11
6.3 A
Valve unit
V5/V6
F12
6.3 A
Reserve
F13
1 AT
Reserve
F14
1 AT
Magnetic valve biopsy chamber
B24
F14
1 AT
Magnetic valve pressure
P31
F15
1 AT
Magnetic valve rotary chamber /
biopsy chamber
P32
F16
1 AT
Magnetic valve vacuum
P30
F17
1 AT
Locking magnet biopsy chamber
B24
F18
1 AT
Reserve 1
F19
1 AT
Reserve 2
F20
1 AT
Pollution degree
2
Overvoltage category
II
Sound pressure
50 dB(A)
Dimensions (mm)
wide 639mm × high 1503mm × deep 635mm
Dimensions (inches)
wide 25.16in × high 59.17in × deep 25.0in
Working height
848mm / 33.39in
Level paraffin containers
1079mm / 42.48in
Overall height, including monitor
1503mm / 59.17in
Weight, empty
190kg
Weight, with operating media
250kg
Note!
Temperatures: All temperatures assume an ambient room temperature of +20°C.