74
Thermo Fisher Scientific
User Guide STP420D/STP420ES
1. Hang all the empty holders that were used in the embedding cycle
in the holder basket located in the rotation chamber or place the
unused holder in the biopsy chamber.
2. Close the lid and press the
Clean
quick start button on the Standby
screen. The ‘Select chamber’ dialog appears.
• A “quick clean” of recently used chamber(s) performed by
default.
3. Press
OK
to confirm.
4. The text box shows the estimated finish time. Press
Yes
to confirm.
Note!
After 50 cycles of quick-cleaning you must perform an extended cleaning
cycle (“standard cleaning”). A reminder will appear on the screen.
3-6 Cleaning Baskets
and replacing
reagents
3-6-1 running a cleaning
cycle
routine operations
Cleaning Baskets and Replacing Reagents