71
Shandon Excelsior ES Operator Guide
c. Highlight
fix2
and select [
REMoVE
].
2. Place the new reagents and ‘tell’ Excelsior ES.
a. Put new bottles of fixative in positions
fix1
and
fix2.
b. Highlight
fix2
and select
[NEw]
.
c. Highlight
fix1
and select
[NEw]
.
This zeroes the Quality Control counter, and removes the
warning triangles from the screen.
Open the lower doors and remove flush reagents in the order
f2, f1, f3.
In the Quality Control screen:
Highlight
1.
f2
and select
[REMoVE].
Repeat for
2.
f1
and
f3
.
Place the new reagents in positions
3.
f3, f1, f2
. In the
Quality Control screen, highlight each reagent in turn,
and select
[NEw].
Excelsior ES rotation is a unique in-process system for
maintaining dehydrants, clearants and infiltrants.
Excelsior ES discards the used reagent from the
•
appropriate containers (used for the first steps in these
processing groups).
Excelsior ES rotates the rest of the reagents.
•
Excelsior ES uses the fresh reagents in the last step for
•
those process groups which have been rotated.
Replacing Fixative
Reagents - Fix1-3
Reagent Rotation
- Overview
Содержание Shandon Excelsior ES
Страница 1: ...Shandon Excelsior ES Operator Guide English A78410120 Issue 3...
Страница 5: ...5 Shandon Excelsior ES Operator Guide...
Страница 12: ...12 Shandon Excelsior ES Operator Guide...
Страница 22: ...22 Shandon Excelsior ES Operator Guide...
Страница 62: ...62 Shandon Excelsior ES Operator Guide...
Страница 101: ...101 Shandon Excelsior ES Operator Guide...
Страница 103: ...103 Shandon Excelsior ES Operator Guide...
Страница 104: ...104 Shandon Excelsior ES Operator Guide...
Страница 112: ...112 Shandon Excelsior ES Operator Guide...
Страница 120: ...120 Shandon Excelsior ES Operator Guide Alcohols from dehydrants list Water Dilute detergent Water Cleaning Agents...