89
Shandon Excelsior ES Operator Guide
Note
1. The filters fit tightly into their slots for efficient air
extraction.
The downdraft filter (potassium permanganate) is at the back
of the instrument, behind the reaction chamber.
1. Remove the filter - slide the metal cover up and remove it.
Slide the filter up and remove the plastic wrappers.
2. Replace the filter, ensuring that the airflow arrow points
away from the instrument.
3. Replace the metal filter cover, locate the bottom of the
cover behind the filter and then push the filter cover down
and towards the front of the instrument. The cover should be
securely positioned on the raised wall.
The optional vent adaptors allow fumes to be extracted into a
fume cupboard or hood, or vented to the outside atmosphere.
DO NOT USE THE VENT ADAPTORS TO EXTRACT
THE FUMES THROUGH THE BUILDING HVAC
SYSTEM OR A COMMON SITE EXTRACTION
SYSTEM.
s
Downdraft
Filter
Cover
Filter
Fitting Optional
Vent Adaptors
Содержание Shandon Excelsior ES
Страница 1: ...Shandon Excelsior ES Operator Guide English A78410120 Issue 3...
Страница 5: ...5 Shandon Excelsior ES Operator Guide...
Страница 12: ...12 Shandon Excelsior ES Operator Guide...
Страница 22: ...22 Shandon Excelsior ES Operator Guide...
Страница 62: ...62 Shandon Excelsior ES Operator Guide...
Страница 101: ...101 Shandon Excelsior ES Operator Guide...
Страница 103: ...103 Shandon Excelsior ES Operator Guide...
Страница 104: ...104 Shandon Excelsior ES Operator Guide...
Страница 112: ...112 Shandon Excelsior ES Operator Guide...
Страница 120: ...120 Shandon Excelsior ES Operator Guide Alcohols from dehydrants list Water Dilute detergent Water Cleaning Agents...