88
Shandon Excelsior ES Operator Guide
BOTH EXTRACTION FILTERS MUST BE FITTED.
DO NOT OPERATE THE INSTRUMENT IF EITHER
EXTRACTION FILTER IS NOT FITTED.
s
FOR CORRECT OPERATION OF THE EXTRACTION
SYSTEM, MAKE SURE THAT THE UPPER AND LOWER
DOORS ARE CLOSED.
s
Excelsior ES is delivered with new filters fitted into the
instrument. Before using the instrument:
Remove the filters
•
Remove the plastic wrappers
•
Replace the filters
•
The two main extraction filters, separated by a removable metal
baffle, are behind the top door. The upper filter uses potassium
permanganate to extract formalin vapours. The lower filter uses
carbon to extract solvent vapours.
To remove the filters,
1. Open the top door and remove the metal baffle plate.
2. Slide out the filters and remove the plastic wrappers.
3. Replace the filters in the correct positions, so that the airflow
arrow on the filter points up.
4. Replace the baffle plate and close the top door.
Fitting Air
Extraction Fitters
Upper Filter
Baffle
Lower Filter
Содержание Shandon Excelsior ES
Страница 1: ...Shandon Excelsior ES Operator Guide English A78410120 Issue 3...
Страница 5: ...5 Shandon Excelsior ES Operator Guide...
Страница 12: ...12 Shandon Excelsior ES Operator Guide...
Страница 22: ...22 Shandon Excelsior ES Operator Guide...
Страница 62: ...62 Shandon Excelsior ES Operator Guide...
Страница 101: ...101 Shandon Excelsior ES Operator Guide...
Страница 103: ...103 Shandon Excelsior ES Operator Guide...
Страница 104: ...104 Shandon Excelsior ES Operator Guide...
Страница 112: ...112 Shandon Excelsior ES Operator Guide...
Страница 120: ...120 Shandon Excelsior ES Operator Guide Alcohols from dehydrants list Water Dilute detergent Water Cleaning Agents...